Load cell for cylinder - ±50 to ±1500 kN - high precision - Fatigue
- High-strength low-profile load cell
- Measurement range: ±50 kN to ±1500 kN (higher on request)
- Certified for fatigue applications
- Static accuracy: ±0.05% to ±0.12% of full scale
- Repeatability: ±0.01%
- Insensitive to transverse forces
- Low deflection
- Temperature sensitivity < 0.0015% / °C
- Tension / compression measurement
- Overload protection: 275%
- Output signal: 2.2 mV/V
- Operating temperature: -55 to 90°C
- Dual or triple bridge option
- See the range of low-profile pancake load cells
Download the product datasheet
Overview
High-performance load cell with flange mounting
The load cells in the 12x8 series differ from the standard 1200 range by their mounting method, using a flange rather than a central threaded hole. This mounting distributes the load across multiple fixing points and allows installation with lower tightening torques. The characteristics of this load cell series are similar to those of the 1200 series: excellent accuracy (0.05%), high robustness, extremely low sensitivity to transverse loads, and high measurement capacities (±50 kN to ±1.5 MN).
They offer a measurement accuracy of < ±0.02% of full scale. The 12X8 models provide very high precision, very low non-linearity, and compensation for non-normal or misaligned forces. They are available for tension/compression or compression-only force measurements.
Design
The selection of strain gauges, the construction of the 12x8 load cell, manufacturing expertise, and calibration ensure static accuracy and linearity of ±0.05% to ±0.12% of full scale depending on the measurement range, very high repeatability (< 0.01%), and long-term stability. The construction is also nearly insensitive to eccentric forces (< 0.01%/mm) and transverse forces (< 0.25%), and highly robust, with overload protection of 275%.

The operating temperature range is -55 to 90°C. Calibration is temperature-compensated between -10 and 45°C.
Specifications
- Measurement: tension / compression
- Bridge supply: 1-20 VDC
- Bridge resistance: 350 Ohm
- Overload protection: 150%
| Model | Ranges | Non-linearity | Hysteresis | Deflection | Output | Weight | Drawing |
|---|---|---|---|---|---|---|---|
| 1228 | ±50 / ±100 / ±140 kN | ±0.05% | ±0.05% | 0.03 mm | 2.2 mV/V | 4.3 kg | Drawing |
| 1238 | ±250 kN | ±0.05% | ±0.05% | 0.05 mm | 2.2 mV/V | 11.8 kg | Drawing |
| 1248 | ±500 kN | ±0.06% | ±0.06% | 0.10 mm | 2.2 mV/V | 32.2 kg | Drawing |
| 1258 | ±1000 kN | ±0.10% | ±0.10% | 0.13 mm | 2.2 mV/V | 46.7 kg | Drawing |
| 1268 | ±1500 kN | ±0.12% | ±0.12% | 0.15 mm | 2.2 mV/V | 92.5 kg | Drawing |
| 1288 | Contact us | ±0.15% | ±0.15% | Contact us | 2.2 mV/V | 204 kg | Drawing |
| 1298 | Contact us | ±0.20% | ±0.20% | Contact us | 2.2 mV/V | 390 kg | Drawing |
Installation
- Bridge supply: 1-20 VDC

Installation is carried out using multiple screws on the base and the sensitive part of the load cell. Specifications for the minimum surface condition and screws to be used are listed in the detailed installation manual.
Signal conditioning and acquisition
The output signal of the 12x8 series load cells is non-amplified. It is possible to obtain a complete measurement chain, wired and configured specifically for the application, by integrating a Display, an Analog signal conditioner/amplifier with voltage or current output, or a Data acquisition system for PC or PLC with included acquisition software. For more information, please contact us.
Applications
- Test benches in aeronautics and automotive
- Test bench instrumentation
- Hydraulic or pneumatic cylinder instrumentation
- Calibration benches
Options and accessories
- Type
- pancake, Press Load
- Precision (max error)
- ≤ 0.1 %
- Nonlinearity (% FS)
- ± 0.05 %
- Hysteresis
- 0.05 % FS
- Nonrepeatability
- 0.01 % FS
- Zero Balance
- 0.0015 %PE/°C
- Sensibilité du gain
- 0.0015 %/°C
- Output Signal
- 2.2 mV/V
- Résistance du pont
- 350 Ω
- Tension d'alimentation
- 10 Vcc
- Measurement direction
- Traction & Compress.
- Surcharge
- 275 %PE
- Housing
- Steel
- Protection Class
- IP40
- Measuring Range (Force)
- 100 - 250 kN, > 250 kN, > 1 MN
- Application
- High Accuracy, OEM, Durability
-
Datasheet_12x8_pmi_EN.pdf
-
Datasheet_12x8_pmi_FR.pdf
-
12X8-Flange-Installation-Rev-B.pdf
-
1228-Drawing.pdf
-
1238-Drawing-1.pdf
-
1248-Drawing-1.pdf
-
1258-Drawing.pdf
-
1268-Drawing.pdf
-
1288-Drawing.pdf
-
CAO 1228-30K
-
CAO 1228-50KN-140KN
-
CAO 1238-55K
-
CAO 1238-250KN
-
CAO 1248-110K
-
CAO 1248-500KN
-
CAO 1258-220K
-
CAO 1258-1000KN
-
CAO 1268-330K
-
CAO 1268-1500KN
-
CAO 1288
-
CAO 1298

- Mechanical protection for the connector(s)
- Standardized outputs
- TEDS
- Connectors: cable, Bayonet or screw-type
- Multiple bridges
- Extended temperature compensation
- Signal conditioning and acquisition electronics
- Cable on request
- Connector
- Instrumentation and acquisition